The Open Door Web Site

 

MPI

Investigation : Starch Calibration Curve

The concentration of a solution can be determined colorimetrically. This is done by comparing it with a series of standardised colours. First a series of solutions is prepared by serial dilution of a stock solution.

In this investigation the concentration of unknown starch solutions is to be determined using iodine solution as an indicator.

A. The Visual Method

Materials

  • 15 test tubes with bungs in a rack

  • 10cm3 graduated pipette and pump

  • 1cm3 graduated pipette and pump

  • Marker pen
  • 50cm3 1% starch solution in a beaker

  • 50cm3 distilled water in a beaker

  • 20cm3 dilute iodine solution

  • 4 unknown starch solutions (A, B, C and D)
  1. Set up  a series of starch solutions as follows:

Starch concentration
/ %

Volume of 1% starch
/ cm3

Volume of distilled water
/ cm3

1 10 0
0.9 9 1
0.8 8 2
0.7 7 3
0.6 6 4
0.5 5 5
0.4 4 6
0.3 3 7
0.2 2 8
0.1 1 9
0 0 10
  1. Label each tube and add 1cm3 iodine solution to each tube replace the rubber bung and turn each tube once.

  2. Take four more test tubes label them and add 10cm3 of solutions A, B, C and D. Add 1cm3 iodine solution to each tube replace the rubber bung and turn once.

  3. Compare the colours of the unknown with the standard solutions. Estimate the concentrations of the unknown solutions.

B. The Colorimetric Method

The colorimeter is a device which will measure the amount of light of a particular wavelength passing through a coloured solution. The wavelength is set by shining light though a filter or by using an LED. The coloured liquid is placed in the path of the light and the amount of light passing through is measured by a photocell.

The colour of the light is often recorded as the wavelength of the light measured in nanometres (nm).

Colours Wavelengths / nm
Red

625 – 665

Orange

605 – 625

Yellow

590 – 605

Green

505 – 590

Blue

436 – 505

Dark Blue (Indigo)

420 – 436

Violet 397 – 420

Materials

TI Graphing Calculator with DataMate program installed
CBL2 interface
Colorimeter sensor

11 cuvettes
Marker pen
Pipette

Starting the DataMate Program and calibrating the sensor

1.   Suck up liquid from a test tube and fill a colorimeter cuvette until it is 1 cm from the top. Wash the pipette, take a sample from the next tube and fill a fresh cuvette. When labelling the cuvettes mark one side only near the top. Handle the cuvettes near the top too.

2.   Use the following steps to start the DataMate program on your calculator:
Press
(for TI-73, 82 and 83 press ), then press the calculator key for the number that precedes DATAMATE. Press . An introductory screen will appear, followed by the main screen.

3.   Plug  the colorimeter probe into channel CH 1 on the CBL2 interface.

4.   Start the DataMate program. Press to reset the program. DataMate will detect the sensor and display the current sensor reading.

Wavelength selection on the colorimeter

The Vernier colorimeter uses four LEDs which shine at a single wavelengths: 430nm, 470nm, 565nm and 635nm. Which colours do these correspond to?

Select the colour of your choice on the colorimeter. An green indicator light will tell you which one you are on.

To decide which colour to select you need to consider the colour of your solution (iodine plus starch is blue so which light does it let through?). Then select a different colour from this.

To calibrate to colorimeter

Make sure the colorimeter has been switched on for about 5 minutes.

Insert a cuvette, filled with distilled and iodine solution in this case, this is called the blank. It will be regulated to show 100% transmittance or 0 absorbance. The clear side of the curvette must face the arrow at the back of the cuvette slot.

Press the CAL button until the red LED begins to flash. Release the CAL button. When the LED stops flashing, the calibration of the colorimeter is finished.

Collecting data

  1. Using the cursor buttons, or  (be patient it’s a bit sluggish!) select MODE, press (scroll up to get to the last item on the menu).

  2. In the SELECT MODE menu press : EVENTS WITH ENTRY.

  3. Press : OK to return to the main screen.

  4. Set the cuvette with the first sample in the colorimeter, take care not to touch the sides where the light will pass.

  5. Select : START to begin data collection. Press to record your measurement. When you are asked to enter value, enter the starch concentration of your first sample. Press again.

  6. Place the next cuvette in the colorimeter and press . Then type in the next starch concentration. Press again and you will find the calculator producing an autoscaled scattergram of the measurements.

  1. Continue to take measurements of all your samples.

  2. You may stop data collection at any time by pressing the key. When you stop data collection  you will see the complete auto-scaled graph.

  3. To store your data, if you are satisfied with it, return to the main screen by pressing .
    Press : TOOLS, then select
    : STORE LATEST DATA RUN. This stores the data in lists. In this case “temperature” in L1 and the colorimeter “absorbance” in L2 with a copy of it in L3.
    To check this press
    : QUIT, then , then and, finally, select : EDIT… You will see a spread sheet with your data in it in L1, L2 (and L3).

Homepage

Physics Homepage

MPI Homepage

MPI Index

>MPI

All About Graphs

Statistics

Listings, Recognitions and Awards

EABJM Teachers' Resource Centre

EABJM Public Web Site

This Site was last
updated on 06/01/08
© The Open Door Team
Any questions or problems regarding this site should be addressed to the webmaster

©  Jacques Gachignard and Paul Billiet 2008